CLINICAL IMAGE QUALITY EVALUATION OF MAMMOGRAPHY FOR BREAST CANCER SCREENING

Autores

  • Flávia Vidal Cabero Universidade Federal de Goiás
  • Ruffo Freitas-Junior Universidade Federal de Goiás
  • Rosângela da Silveira Corrêa National Nuclear Energy Commission/Center for Nuclear Sciences in the Midwest
  • Danielle Cristina Netto Rodrigues Universidade Federal de Goiás
  • Leonardo Ribeiro Soares Universidade Federal de Goiás
  • Paola Ferreira de Freitas Hospital Pérola Byington

Palavras-chave:

Screening, Mammography, Clinical Quality

Resumo

Objectives: To evaluate mammographic image quality in a real clinical practice scenario for breast cancer screening. Materials and Methods: Observational prospective study where images from digital mammograms from Diagnostic Service (DS) in the state of Goiás in 2019 were analyzed. A specific protocol was created based on evaluation criteria of the Brazilian College of Radiology, European Guidelines, and American College of Radiology. For each variable, score 1 was attributed to conformity and zero for nonconformity. Logistic regression model was utilized using the following independent variables: location (city vs. country), Public health system – Sistema Unico de Saude (SUS) (public vs. private), number of monthly examinations (≤300 vs. >300), device manufacturing year (≤2011 vs. >2011), and breast density (≤75% vs. >75% of the parenchyma). Results: Of the 163 fully functioning mammograms, 151 (92.6%) were eligible, providing 1,024 images. We evaluated 12,032 items, of which 4,096 were craniocaudal projections, 4,608 mediolateral oblique projections, and 3,328 equipment related. On the clinical image analysis as to the positioning of the patient, the higher conformity for symmetry parameters, in both projections (>90%), was observed. The conformity rate among the other parameters varied from 18.6% to 100%. In the multivariable analysis, it was observed that only the variables monthly examinations (OR 3.44; 95%CI 1.67–7.09; p=0.0008) and mammogram device manufacturing year (OR 2.46; 95%CI 1.02–5.95; p=0.04) were related to a higher conformity rate. After the percentage consolidation conformity rate per DS, as to the final clinical mammography quality, no DS presented desirable conformity (>90%), 20 DS obtained acceptable conformity (between 70% and 89%), and 10 DS presented conformity below 70%. Conclusion: Conformity rate of mammographic examination is extremely low and varies according to the multiple parameters analyzed. Mammographies performed at centers with less productivity (≤300/monthly) and with newer devices (>2,011) presented higher chances of conformity at the clinical imaging evaluation.

Downloads

Não há dados estatísticos.

Biografia do Autor

Flávia Vidal Cabero, Universidade Federal de Goiás

Brazilian Breast Cancer Research Network, Mastology Program, Hospital das Clínicas, Universidade Federal de Goiás

Ruffo Freitas-Junior, Universidade Federal de Goiás

Faculty of Medicine, Hospital das Clínicas, Universidade Federal de Goiás

Paola Ferreira de Freitas, Hospital Pérola Byington

Resident of the Mastology Program of the Women’s Health Reference Center , Hospital Pérola Byington

Downloads

Publicado

2021-10-14

Como Citar

Cabero, F. V., Freitas-Junior, R., Corrêa, R. da S., Netto Rodrigues, D. C., Soares, L. R., & Freitas, P. F. de. (2021). CLINICAL IMAGE QUALITY EVALUATION OF MAMMOGRAPHY FOR BREAST CANCER SCREENING. Mastology, 31, 53. Recuperado de https://revistamastology.emnuvens.com.br/revista/article/view/1026

Edição

Seção

E-poster